Finding hidden tumors by the light they give off

Where
Xstrahl
2012–2020
Founding engineer → CTO → COO

Some tumors are invisible on an X-ray scan because they look just like the tissue around them. Tumor cells can be made to glow faintly, the way a firefly does, a property called bioluminescence. This device sees that glow from several sides and works out where the tumor is, so the machine can aim at it.

I led turning a Johns Hopkins research prototype into a desktop-computer-sized product, and we brought it to market in 2015.

The unit, from the sideinstrument chamberimaging chamber (dark)bed slides inM1M2M3rotation axisthe 3 mirrors turn 360°around the subjectfixed mirrorfilter wheellenscooledcameraLooking along the bedM1 orbits the subject
The bed slides into a dark imaging chamber. A set of 3 mirrors turns 360° around the subject and passes the glow along the rotation axis to a fixed mirror, which turns it down through a filter wheel and lens to a cooled camera. Matched to the outline from the CT scan, software works back to where the light started.

Tumor cells that make their own light

Light from inside tissue spreads out before it reaches the surface, so a single picture shows only a blur. The bed slides into a dark imaging chamber, where a set of mirrors turns 360° around it and carries its glow, from whichever side it faces, to a fixed mirror, a filter wheel and a single cooled camera. Matched to the outline from the CT scan, software works back to where the light started, and that point becomes the target.

From a lab prototype to a desktop product

It grew out of a National Institutes of Health partnership (2011–2015) between Johns Hopkins and us. The prototype was an optical assembly under a light-tight dome that had to be docked onto the irradiator by hand, built around a large research camera. We made it a desktop-sized, light-tight unit that works on its own on a bench, with a removable bed that carries the subject straight to the irradiator, a fourth mirror to keep it compact, a new camera (the Andor iKon-M 934) and CE marking.